ECS-451 DDD

Implantable dual-chamber pacemaker DDD with telemetry

Implantable dual-chamber pacemaker DDD with telemetry

Parameters & values

Mode DDD,VVI,DVI,DOO,VOO,DAD
Rate 30, 35, 40, 45, 50, 55, 60, 65, 70, 75,80, 85, 90, 95, 100,105, 110, 115, 120,125, 130, 135,140, 145, 150 ppm
Pulse amplitude: Atrium 0.5; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0; 5.5; 6.0; 6.5; 7.0; 7.5 V
Ventricle 0.5; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0; 5.5; 6.0; 6.5; 7.0; 7.5 V
Pulse width: Atrium 0.25; 0.5; 0.75; 1.0 ms
Ventricle 0.25; 0.5; 0.75; 1.0 ms
Sensitivity: Atrium 0.6; 0.9; 1.3; 1.9; 2.4; 3.5; 5.0; 7.5 mV
Ventricle 0.9; 1.3; 1.9; 2.7; 3.5; 5.0; 7.0; 11.0 mV
Refractory period: Atrium 250, 312, 437, ∞ (asynchronous mode)
Ventricle 250, 312, 437, ∞ (asynchronous mode)
A-V delay 0, 62, 125, 187 ms
Hysteresis 0, 125, 250, 375 ms
Magnet rate 100 ppm
VARIO stimulation threshold test ON / OFF
Replacement indication magnet rate decrease down to 80 ppm
Expected service life 7 years (100% pacing at standard program)
Nominal service life up to 10 years (depending on patient’s needs and mode of operation)
Lead polarity: pacing/sensing unipolar / unipolar
Lead connection IS–1 UNI
Dimensions 60×47×10 mm
Mass not over 45 g
Telemetry measurement of parameters

Standard program

  ATRIUM VENTRICLE
Rate 60 ppm 60 ppm
Pulse amplitude 4 V 4 V
Pulse width 0.5 ms 0.5 ms
Sensitivity 0.9 mV 5 mV
Refractory period 312 ms 312 ms
Hysteresis 0 ms 125 ms
VARIO test OFF OFF
Lead polarity uni /uni uni /uni

SHIPPED SETTINGS the pacemaker is delivered with preset DDD-mode

All data at 37 °C, 510 Ohms